跳转至内容
Merck
CN

G006153

LC/MS (TOF) Analysis of Opioids and Opiate-Dependence Management Drugs on Ascentis® Express HILIC

application for HPLC

登录查看公司和协议定价


About This Item

技术

HPLC: suitable

试验参数

column: Ascentis Express HILIC, 5 cm x 2.1 mm I.D., 2.7 μm particles (53934-U)
mobile phase: 2 mM ammonium formate in 90:10 (v:v) acetonitrile:water
flow rate: 0.3 mL/min
column temp.: 55 °C
detector: TOF/MS
injection: 2 μL
sample: 400 ng/mL in acetonitrile

适用性

application for HPLC

应用

clinical testing
forensics and toxicology

分析说明

The rapid analysis by HPLC with TOF-MS detection of seven opioids is shown here using an Ascentis Express HILIC column. Cerilliant CRMs were used to ensure reliable MS identification and quantification.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Oxycodone solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    6-Acetylcodeine solution, 1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    6-Acetylmorphine, 1.0 mg/mL solution, 1 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    Codeine solution, 1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Hydrocodone solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Morphine solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Hydromorphone solution, 1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Choose from one of the most recent versions:

分析证书(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门