跳转至内容
Merck
CN

G005791

LC/MS Analysis of L-Methionine Sulfoximine and Sulfone on Ascentis® Express OH5

application for HPLC

登录查看公司和协议定价

技术

HPLC: suitable

试验参数

column: Ascentis Express OH5, 10 cm x 3.0 mm I.D., 2.7 μm particles (53769-U)
mobile phase: [A] acetonitrile; [B] water with 0.1% formic acid, pH 3.5 with ammonium formate; (75:25, A:B)
flow rate: 0.4 mL/min
pressure: 1350 ppi (93 bar)
column temp.: 25 °C
detector: MS, ESI (+), TIC, m/z 150-300
injection: 5 μL

适用性

application for HPLC

应用

chemicals and industrial polymers

分析说明

L-methionine sulfoximine and L-buthionine sulfoximine are used in are used to prevent additional enzyme activity in Chinese hampster ovary (CHO) cell lines with supplemental Glutamine Synthase (GS). It is of interest to be able to monitor both purity of such molecules in order to control addition to cell media as well as for the assay of parent molecules during use. The analytes are highly polar and thus should be amenable to HILIC-LC-MS analysis. This first installment investigates the ability of Ascentis Express HILIC and Ascentis Express OH5 stationary phases for the separation of the methionine analog and related compounds as well as methionine and buthionine separations.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    L-Methionine sulfone
    L-Methionine sulfoximine, suitable for cell culture

标准

产品编号
说明
价格

分析证书(COA)

输入产品批号来搜索 分析证书(COA) 。批号可以在产品标签上"批“ (Lot或Batch)字后找到。

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门