G005723

LC/MS Analysis of Polypeptide Drugs on Ascentis® Express RP-Amide

application for HPLC

登录查看公司和协议定价


About This Item

技术

HPLC: suitable

试验参数

column: Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)
mobile phase: [A] 10 mM ammonium acetate in water, pH 4.2; [B] 10 mM ammonium acetate in 90:10 (v/v) acetonitrile:water, pH 4.7
gradient: held at 20% B for 0.1 min; 20 to 80% B in 2.9 min; 80 to 20% B in 0.01 min; held at 20% B for 0.99 min
flow rate: 0.3 mL/min
column temp.: 35 °C
detector: TOF, ESI(+), m/z 100-2500
injection: 3 μL
sample: 1 μL/mL in 10mM ammonium formate, 80:20, acetonitrile:water

适用性

application for HPLC

应用

life science and biopharma

分析说明

This application demonstrates the suitability of the Ascentis Express RP Amide for the analysis of polypeptide drugs.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Oxytocin
    [deamino-Cys1, D-Arg8]-Vasopressin acetate salt hydrate