跳转至内容

Dear Customer:

The current international situation is complex and volatile, and uncertain tariff policies may potentially impact our product prices. Given these uncertainties, we value your understanding regarding order-related matters.

If you decide to place an order during this period, we reserve the right to adjust the price based on the evolving situation. We understand that market changes may cause inconvenience. We will negotiate with you if there’s a significant price fluctuation due to tariff policy changes before the order’s actual delivery, and in such cases we may adjust or cancel the order as necessary.

Merck
CN
HomeqPCRMultiplex qPCR Protocol

Multiplex qPCR Protocol

PCR Technologies Protocols Table of Contents

Quantitative PCR Protocols

Although quantitative PCR uses the same basic concept as traditional PCR, the reactions differ in that the amplicons are generally smaller and are detected indirectly using an additional dye or labeled probe or primer.


In the PCR Technologies Guide, the requisite components and quality control requirements for qPCR experiments were described in detail. With those in mind, the following is a protocol that can be used as a basic template for qPCR incorporating up to four detection probes. In these reactions, primers and probe are included at a final concentration of 200 nM and are run using LuminoCt® ReadyMix. However, multiplex experiments require optimization and it is advisable to test the assay combinations by adding each to the multiplex sequentially. A detailed assay optimization protocol for each single assay is described in Primer Concentration Optimization and Primer Optimization Using Temperature Gradient.

Equipment

  • Quantitative PCR instrument
  • Microcentrifuge
  • Laminar flow hood for PCR set up (optional)

Supplies

  • DNA/cDNA template: cDNA reaction diluted 1:10 to detect a medium to highly expressed targets or 1:2 to 1:5 for rare transcripts. 10 ng to 100 ng gDNA.
  • LuminoCt® ReadyMix™ (L6669).
  • PCR grade water: PCR grade water (W1754 or W4502) as 20 mL aliquots; freeze; use a fresh aliquot for each reaction.
  • Forward and reverse primers concentration stocks (100 μM working stocks are suitable for use in multiplex reactions).
  • Specific target detection probes (see Chapter 6) concentration stocks (100 μM working stocks are suitable for use in multiplex reactions).
    • Custom oligos can be ordered here.
  • Sterile filter pipette tips
  • Sterile 1.5 mL screw-top microcentrifuge tubes (CLS430909)
  • PCR tubes and plates, select one to match desired format:
    • Individual thin-walled 200 μL PCR tubes (Z374873 or P3114)
    • Plates
        - 96-well plates (Z374903)
        - 384-well plates (Z374911)
    • Plate seals
        - ThermalSeal RTS™ Sealing Films (Z734438)
        - ThermalSeal RT2RR™ Film (Z722553)

Method

  1. Defrost all reaction components on ice, taking care to protect the probes from exposure to light.

  2. Calculate the number of reactions required to enable samples and controls to be run in duplicate. Include two
    No Template Controls (NTC).

  3. Prepare a probe blend and a primer blend using Tables P6‑13A and 13B as a guide. For reactions
           requiring 2 or 3 probes, adjust to the total volume (with PCR grade water). After optimizing primer
           concentration, adjust primer concentrations/volumes accordingly. The volumes stated in Tables P6‑13A
           and 13B are sufficient to run 250 reactions, scale up accordingly for more reactions.
Probe (100 μM)Volume (μL)
Target 110
Target 210
Target 310
Target 4
10
PCR grade water60
Table P6-13AVolume Of Stock Probe to Blend for 200 nM Final Concentration in Each Reaction. Amend volumes in the blend following optimization.
Forward Primer
(100 μM)
Volume
(μL)
Reverse Primer
(100 μM)
Volume
(μL)
Target 110Target 110
Target 2
10Target 2
10
Target 3
10Target 3
10
Target 4
10Target 4
10
PCR grade water  20
Table P6-13BVolume of Stock Primers to Blend for 200 nM Final Concentration in Each Reaction. Amend volumes in the blend following optimization.

4.    Prepare a master mix for all reactions according to Table P6‑14 (calculate volumes for each reaction and add 10% to
         allow for pipetting error). Mix well, avoiding bubbles.

ReagentVolume (μL) per
Single 20 μL Reaction
2× LuminoCt qPCR ReadyMix10
Probe Mix (Table P6-13A)0.4
Primer mix (Table P6-13B)0.4
PCR grade water4.2
Table P6-14Master Mix for Mulitplex Probe qPCR.

5.    Add 5 μL of water to the NTC reaction tubes.

6.    Add 5 μL of cDNA/gDNA solution to the appropriate tubes/wells.

7.    Aliquot 15 μL template master mix remaining from step 4 into the PCR tubes.

8.    Cap tubes or seal the PCR plate and label (according to instrument requirements). (Make sure the labeling
       does not obscure instrument excitation/detection light path.)

9.    Centrifuge briefly and visually check that all tubes/wells contain sample at the bottom at the correct volume.

10.   Run samples according to the two-step protocol (Table P6‑15), repeating steps 1–2 through 40 cycles.
        (Note: These conditions are specific for FAST cycling protocols).

11.   For reactions containing Scorpions® Probes or Molecular Beacons, a three-step protocol (Table P6‑16) may
        result in more efficient/sensitive detection. When adopting this protocol, the annealing temperature of step 2
        can be optimized. Cycle steps 1–3 through 40 cycles.

FAST Cycling ConditionsTemp (°C)Time (sec)
Initial Hot Start/denaturation9530
Steps 1–2 are repeated through 40 cycles
Step 1955
Step 2
6030
Table P6-15Fast PCR Cycling Conditions.
3-Step Cycling Conditions (for Use with Scorpions®
Probes or Molecular Beacons)
Temp (°C)Time (sec)
Initial denaturation9530
Steps 1–3 are repeated through 40 cycles
Step 1 Denature955
Step 2 Anneal (select optimized Ta)5815
Step 3 Extend7210
Table P6-16Three-step PCR Cycling Conditions for use with Scorpions® Probes or Molecular Beacons.
登录以继续。