跳转至内容
Merck
CN

HPLC Analysis of Benalaxyl Enantiomers on Astec CYCLOBOND I 2000 HP-RSP in Reversed Phase Mode

HPLC Analysis of Benalaxyl Enantiomers on Astec CYCLOBOND™ I 2000 HP-RSP in Reversed Phase Mode application for HPLC

材料

分析色谱柱

产品编号
说明
价格

Astec® CYCLOBOND I 2000 HP-RSP手性HPLC色谱柱

5 μm particle size, L × I.D. 25 cm × 4.6 mm

标准

产品编号
说明
价格

苯霜灵

PESTANAL®, analytical standard

流动相组分

产品编号
说明
价格

乙酸铵

LiChropur, eluent additive for LC-MS

乙腈

suitable for HPLC, gradient grade, ≥99.9%

CONDITIONS

column

Astec CYCLOBOND I 2000 HP-RSP, 25 cm x 4.6 mm I.D., 5 μm particles (24024AST)

mobile phase

[A] 10 mM ammonium acetate, pH 4.0 with acetic acid; [B] acetonitrile; (70:30, A:B)

flow rate

1 mL/min

pressure

1550 psi (107 bar)

column temp.

25 °C

detector

UV, 220 nm

injection

10 μL

sample

250 μg/mL in 75:25, water:methanol

说明

分析说明

Benalaxyl is a fungicide used to protect a variety of crops, including potatoes and grapes. It is a chiral compound, comprising two enantiomers. Enantioselective influences of chiral pesticides, like Benalaxyl, and their metabolites on ecosystems is an area of great importance. The possibility that only one enantiomer has the wanted properties and/or the other form an unwanted effect may exist. An analytical separation of the enantiomers is required to perform studies in these areas. The Astec CYCLOBOND I 2000 HP-RSP HPLC column resolves the enantiomers of Benalaxyl under reversed-phase and LC/MS-compatible conditions.

法律信息

null