GC Analysis of Pesticides in Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Pesticides in Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

SPME Fiber Assembly

85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, white hub

相关产品

产品编号
说明
价格

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

sample/matrix

white wine spiked with 50 ppb pesticides

SPME fiber

85 μm polyacrylate (57304)

extraction

immersion, room temp (30 min)

desorption process

250 °C for 5 min

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

60 °C (1 min), 15 °C/min to 100 °C, 7 °C/min to 300 °C (1 min)

MSD interface

325 °C

scan range

SIM

carrier gas

helium, 0.7 mL/min, constant

liner

0.75 mm I.D. SPME liner