技术
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
试验参数
sample preparation: Dispersive (QuEChERS)
sample/matrix: 10 gm homogenized beef kidney spiked at 50 ng/g with pesticides
extraction tube: Supel Que Acetate (AC) Tube (55234-U)
extraction process: add 10 mL of acetonitrile; shake for 1 minute; add contents of Supel Que acetate extraction tube (55234-U); shake immediately for 1 minute; centrifuge at 3200 rpm for 5 minutes; (Sample Pretreatment: Homogenize 10 gm beef kidney and spike with pesticides at 50 ng/g.)
clean-up tube: Supel QuE Z-Sep+ tube (55296-U)
clean-up process: transfer 3 mL of the acetonitrile layer into a Supel QuE Z-Sep+ cleanup tube (55296-U); shake for 1 minute; centrifuge at 3400 rpm for 3 minutes; draw off supernatant for LVI-GCMS analysis
column: F.S. CAP. SLB-5MS 20 m x 0.18 mm I.D., 0.36 μm (28576-U)
oven: 70 °C ( 0.5 min), 25 °C/min to 125 °C, 10 °C/min to 200 °C, 5 °C/min to 300 °C/min (1 min)
inj. temp.: programmed, 60 °C (0.28 min), 600 °C/min to 325 °C (5 min)
detector: MSD
MSD interface: 325 °C
carrier gas: helium, 1 mL/min constant flow
injection: 10 μL LVI, PTV solvent vent, rapid injection speed
liner: 4 mm ID FocusLiner with taper
sample: Beef Kidney extract
适用性