技术
HPLC: suitable
试验参数
sample/matrix: rabbit plasma, unfiltered K2-EDTA spiked with Spice cannabinoids (5 ng/mL each)
SPE well plate: HybridSPE-Phospholipid, 96-well plate, 50 mg bed wt., 2 mL well vol (575656-U)
sample addition: to each well add 100 μL plasma, followed by a 300 μL of 1% formic acid in acetonitrile, agitate on orbital shaker for 2 minutes
elution: attach collection plate and apply vacuum at 10” Hg for 4 minutes
column: Ascentis Express F5, 5 cm x 2.1 mm I.D., 2.7 μm particles (53567-U)
mobile phase: [A] 10 mM ammonium formate in water, pH 6.8 (unadjusted); [B] acetonitrile; (50:50; A:B)
flow rate: 0.3 mL/min
pressure: 1296 psi (89 bar)
column temp.: 35 °C
detector: MS, ESI(+), MRM, m/z 344/155 (JWH-073 metabolite), 385/155 (JWH-200), 336/121 (JWH-250), 328/155 (JWH-073), and 342/155 (JWH-018)
injection: 2 μL
适用性
application for HPLC
应用
forensics and toxicology
分析说明
法律信息
分析物
分析色谱柱
标准
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门