跳转至内容
Merck
CN
所有图片(1)

文件

1.09088

Millipore

穿孔盖

400 hole x 0.7mm, Aluminium, for MAS-100 Eco®

登录查看公司和协议定价

别名:
Collection head, Sieve head, Perforated Lid for MAS-100NT, 400 hole x 0.7mm
UNSPSC代码:
41103403
NACRES:
NA.74

物料

aluminum lid

Agency

non-compendial

特点

no. of holes 400 autoclavable

包装

pkg of 1 unit

参数

11 m/s impaction speed (at flow rate of 100 L/min)

孔直径

0.7 mm

应用

air monitoring
food and beverages

相容性

for use with 90 mm Petri Dishes
for use with MAS-100 Eco®

储存温度

no temp limit

正在寻找类似产品? 访问 产品对比指南

一般描述

The Perforated 400 x 0.7 mm hole lid is made from anodized aluminum and can be used with all portable MAS microbial air samplers, especially designed for the MAS-100 Eco. The sieve of the perforated lid is at the core of the air sampler. Each hole must have an identical diameter and length to accelerate the aspirated air correctly. The 3mm thickness of the perforated sieve guarantee a well-defined impaction jet. The impaction speed defines the physical sampling efficiency. The cut-off value D50 (the particle size at which 50% of the particles impact in the agar due to inertia and 50% are exhausted with the airstream) must be identical between holes and between lids. That is why intense quality control of the sieves is performed. The sieve pattern with 400 holes of a diameter of 0.7 mm provides an impaction speed of approximately 11 m/s at a flow rate of 100 l/min. This impaction speed allows a high physical efficiency, meaning a sampling efficiency of more the 50% of small particles at a size of less than 1.62 1μm (D50) while maintaining the viability of microorganisms during impaction on the agar surface. The aluminium perforated lid can be autoclaved and disinfected with all disinfectants compatible anodized aluminium in order to avoid false positive results in high grade cleanrooms. This permit using a new lid for each sampling position or production batch or using lids while others are in the sterilization process and inaccessible. The sampling efficiency for an instrument must remain identical, irrespective of the lid used. Our perforated lids allow the exchange of several lids with identical sieve pattern on the MAS air sampler without the need of re-calibration.

应用

The Perforated Lid with 400 x 0.7 mm hole sieve is used for microbial monitoring of ambient air with portable MAS-100 microbial air samplers. Especially MAS-100 Eco. Microbial air sampling devices are extensively used in Pharmaceutical and Personal care industries as well as Food, Beverage and Environmental applications. Additionally, a regulation for compounding pharmacies, USP797, that requires active air sampling in the level 5 and above areas.

联系

Replaces: 1.09088.0002

法律信息

MAS-100 Eco is a registered trademark of MBV AG

储存分类代码

10-13 - German Storage Class 10 to 13


分析证书(COA)

输入产品批号来搜索 分析证书(COA) 。批号可以在产品标签上"批“ (Lot或Batch)字后找到。

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

商品

因制药环境中的卫生标准与成品的微生物安全性密切相关,因此活性微生物空气采样解决方案的精确性、可靠性和可重复性不容忽视。默克面向制药行业的微生物空气监测:MAS®-100系列,默克微生物空气采样监控系统简单易用。

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门