跳转至内容
Merck
CN

HPLC Analysis of PAHs in Water on SUPELCOSIL LC-PAH after SPME using 100 μm PDMS Fiber

HPLC Analysis of PAHs in Water on SUPELCOSIL™ LC-PAH after SPME using 100 μm PDMS Fiber application for SPME, application for HPLC

材料

SPME 光纤

产品编号
说明
价格

SPME 纤维组件聚二甲基硅氧烷 (PDMS)

df 100 μm(PDMS, for use with autosampler, needle size 24 ga

分析色谱柱

产品编号
说明
价格

SUPELCOSIL LC-PAH HPLC 色谱柱

5 μm particle size, L × I.D. 15 cm × 4.6 mm

CONDITIONS

sample/matrix

5 μL PAH mix (Cat. No. 4-8743) in 5 mL water

SPME fiber

polydimethylsiloxane, 100 μm (57301)

extraction

immersion, 30 min (rapid stirring)

desorption process

static, 200 μL acetonitrile:water, 40:60, 2 min

column

SUPELCOSIL LC-PAH, 15 cm × 4.6 mm I.D., 5 μm particles (58318)

gradient

0 min - 50% A; 5 min - 50% A; 30 min - 100% A; 45 min - 100% A

flow rate

0-2 min: 0.2 mL/min 2-45 min: 1.0 mL/min

mobile phase

[A] acetonitrile:[B] water

detector

UV, 254 nm