跳转至内容
Merck
CN

GC Analysis of QA Test Mix on SLB®-5ms, after Extended Hold at 360 °C

GC Analysis of QA Test Mix on SLB®-5ms, after Extended Hold at 360 °C suitable for GC

材料

分析色谱柱

产品编号
说明
价格

SLB ®-5 ms 毛细管 GC 色谱柱

L × I.D. 30 m × 0.25 mm, df 1.00 μm

标准

产品编号
说明
价格

GC 非极性柱测试混标

certified reference material, 500 μg/mL in dichloromethane

CONDITIONS

column

SLB-5ms, 30 m x 0.25 mm I.D., 1.0 μm (28476-U)

oven

145 °C (20 min), 15 °C/min to 325 °C (15 min), 15 °C/min to 340 °C (15 min), 15 °C/min to 360 °C (4 hours)

inj. temp.

250 °C

detector

FID, 360 °C

carrier gas

helium, 30 cm/sec. @ 145 °C

injection

1.0 μL 100:1 split

liner

4 mm I.D. split, cup design

sample

Non-polar Column Test Mix (47300-U), 500 μg/mL in methylene chloride

说明

分析说明

Column evaluation test mix, before and after column held for 76 hours at 360 °C, 20 degrees above the published maximum isothermal temperature limit.

法律信息