跳转至内容
Merck
CN

GC Analysis of Aromatics in Reformulated Gasoline on SLB®-IL111

GC Analysis of Aromatics in Reformulated Gasoline on SLB®-IL111 suitable for GC

材料

分析色谱柱

产品编号
说明
价格

SLB®-IL111毛细管GC色谱柱

L × I.D. 60 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

SLB-IL111, 60 m x 0.25 mm I.D., 0.20 μm (28928-U)

oven

50 °C (3 min), 15 °C/min to 265 °C (5 min)

inj. temp.

250 °C

detector

FID, 265 °C

carrier gas

helium, 30 cm/sec

injection

0.5 μL, 200:1 split

liner

4 mm I.D., splitless type, single taper wool packed FocusLiner design

sample

reformulated unleaded gasoline

说明

分析说明

This application illustrates the analysis of a sample of reformulated gasoline on a 60 m SLB-IL111 ionic liquid column. Due to the extremely polar nature of the phase, the aliphatics elute prior to the aromatics. The high level of ethanol in reformulated gasoline can pose a problem in the detection and quantification of benzene. The SLB-IL111 provided separation between these two compounds, with a starting temperature of 50 °C providing the best resolution. The SLB-IL111 has a maximum operating temperature of 270 °C, which is extremely high for a phase of this polarity. The high maximum temperature was utilized in this analysis to elute the napthalenes in < 15 minutes.

法律信息

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany