跳转至内容
Merck
CN
  • Development and validation of an ion-pair reversed-phase high-performance liquid chromatographic method for the separation of phosphonodipeptides.

Development and validation of an ion-pair reversed-phase high-performance liquid chromatographic method for the separation of phosphonodipeptides.

Journal of chromatography. B, Biomedical sciences and applications (1997-04-11)
N Surendran, C J Stankovic, B H Stewart
摘要

The objective of this research was to develop a rapid, sensitive and reliable method for the separation of phosphonodipeptide prodrugs and parent compounds to facilitate the evaluation of cell permeation using in vitro cell culture models. Separation was accomplished isocratically within 10.0 min using a C18 (150x4.6 mm I.D., 3 microm) reversed-phase column. The mobile phase consisted of 5 mM tetrahexyl ammonium (ion-pair reagent) in 0.02 M phosphate buffer pH 6.5-acetonitrile (48.5:51.5, v/v). The flow-rate was 1.1 ml/min with detection at 221 nm. The standard curves were linear (r2>0.999) over the concentration range 1-100 microM. The method was reliable and reproducible, with the limit of quantitation being 1 microM (25 ng on column).

材料
货号
品牌
产品描述

Sigma-Aldrich
四己基溴化铵, 99%
Sigma-Aldrich
硫酸氢四己铵, ≥98.0% (T)
Supelco
硫酸氢四己铵, suitable for ion pair chromatography, LiChropur, ≥99.0% (T)
Sigma-Aldrich
四己基氯化铵, 96%
Sigma-Aldrich
四己基氢氧化铵 溶液, ~40% in H2O (T)
Supelco
四己基溴化铵, suitable for ion pair chromatography, LiChropur, ≥99.0% (AT)