技术

HPLC: suitable

试验参数

column: Ascentis Express OH-5, 15 cm x 4.6 mm, I.D., 2.7 μm particles (53778-U)
mobile phase: [A] 0.005 M ammonium formate, pH 3.15 with concentrated TFA [B] acetonitrile (30:70, A:B)
flow rate: 0.5 mL/min
pressure: 841 psi (58 bar)
column temp.: 35 °C
detector: ELSD (detector other: Nebulizer 60º C, Evaporator 80º C, Gas flow 1.60 SLM (Standard Liters/ Minute), LED 100%, Smoothing 5.0 S, PMT ( Photomultiplier) Gain 1.0)
injection: 10 μL
sample: 0.5 mg/mL each prepared in 50:50 Water:Acetonitrile

适用性

application for HPLC

应用

food and beverages

分析说明

This application shows the analysis of sugars by HILIC Chromatography with ELSD detection. The purpose of the work shown here was to develop a HILIC method that could quickly separate mono- and di-saccharides with good sensitivity and reproducibility utilizing the Ascentis Express OH-5 column and an Evaporative Light Scattering Detecctor (ELSD) detector.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Dextrose, Pharmaceutical Secondary Standard; Certified Reference Material
    Sucrose, Pharmaceutical Secondary Standard; Certified Reference Material
    Maltitol, Pharmaceutical Secondary Standard; Certified Reference Material

Choose from one of the most recent versions:

分析证书(COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact 客户支持

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门