技术
HPLC: suitable
试验参数
column: Ascentis Express C18, 10 cm x 3.0 mm I.D., 2.7 μm particles (53814-U)
mobile phase: [A] 20 mM ammonium acetate, pH 6.4 unadjusted; [B] acetonitrile
gradient: 20 to 28% B in 11 min; 28 to 65% B in 5 min; held at 65% B for 4 min
flow rate: 0.6 mL/min
column temp.: 46 °C
detector: UV, 240 nm
injection: 5 μL
sample: 10-25 μg/mL in 10:90, water: acetonitrile
适用性
application for HPLC
应用
agriculture
分析说明
This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of 17 triazine pesticides.
法律信息
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
说明
Atrazine-desethyl
Metoxuron
Hexazinone
Simazine
Cyanazine
Methabenzthiazuron
Atrazine
Monolinuron
Diuron
Isoproturon
Metobromuron
Metazachlor
Sebuthylazin
Linuron
Metolachlor
查看所有结果 (15)
分析色谱柱
标准
产品编号
说明
价格
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门