技术
gas chromatography (GC): suitable
试验参数
column: Equity-1, 60 m x 0.25 mm I.D., 0.25 μm (28047-U)
oven: 50 °C (3 min), 15 °C/min to 265 °C (5 min)
inj. temp.: 250 °C
detector: FID, 265 °C
carrier gas: helium, 30 cm/sec
injection: 0.5 μL, 200:1 split
liner: 4 mm I.D., splitless type, single taper wool packed FocusLiner™ design
sample: reformulated unleaded gasoline
应用
petroleum
分析说明
This application illustrates the analysis of a sample of reformulated gasoline on the Equity-1 column. Due to the nonpolar nature of the phase, there is a great deal of overlap in the elution ranges of the aliphatic and aromatic portions of this highly complex sample. Benzene is not baseline resolved even on a 60 m column. For this reason, this phase could not be used by itself for the analysis of benzene in this sample.
法律信息
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
分析物
说明
Ethanol denatured
Pentane
Hexane
Benzene
Heptane
Toluene
Octane
Ethylbenzene
p-Xylene
m-Xylene
o-Xylene
Nonane
1,2,4-Trimethylbenzene
Decane
1,2,3-Trimethylbenzene
Undecane
Durene
Naphthalene
2-Methylnaphthalene solution
1-Methylnaphthalene
查看所有结果 (20)
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门