跳转至内容
Merck
CN

796-0376

OSHA Method 42/47 and ASTM® D5836: HPLC Analysis of Isocyanates on SUPELCOSIL LC-8 after Collection/Desorption using ORBO-80

application for air monitoring, application for HPLC

登录查看公司和协议定价

技术

HPLC: suitable

试验参数

sample/matrix: calibration standard: isocyanate derivatives, 5 μg/mL in acetonitrile blank: 1-(2-pyridyl)piperazine (1-2PP) coated glass fiber filter desorbed in acetonitrile/DMSO, 9:1
adsorbent tube: ORBO-80 (20811)
column: SUPELCOSIL LC-8, 25 cm × 4.6 mm I.D., 5 μm particles (58297)
mobile phase: 0.05 M ammonium acetate in water:acetonitrile, 70:30 (pH 6-6.2 with acetic acid)
flow rate: linear gradient, 0.7 mL/min to 2 mL/min in 15 min
detector: UV, 254 nm
injection: 10 μL

适用性

application for HPLC
application for air monitoring

应用

air monitoring
industrial hygiene

法律信息

ASTM is a registered trademark of American Society for Testing and Materials
ORBO is a trademark of Sigma-Aldrich Co. LLC
SUPELCOSIL is a trademark of Sigma-Aldrich Co. LLC

分析物

说明

    1-(2-Pyridyl)piperazine
    Hexamethylene diisocyanate
    2,6-Tolylene diisocyanate
    2,4-Tolylene diisocyanate

分析证书(COA)

输入产品批号来搜索 分析证书(COA) 。批号可以在产品标签上"批“ (Lot或Batch)字后找到。

已有该产品?

在文件库中查找您最近购买产品的文档。

访问文档库

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门