技术
gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable
试验参数
sample/matrix: water spiked with PAHs and phthalates at 50 ppb
SPME fiber: polydimethylsiloxane, 7 μm (57302)
extraction: direct immersion in water, 15 min (rapid stirring)
column: 5% Phenyl Polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 60 °C (3 min), 10 °C/min to 320 °C
inj. temp.: 280 °C
detector: MS, scan range m/z = 45-465 at 0.6 sec/scan
carrier gas: helium, 40 cm/sec at 60 °C
适用性
application for SPME
应用
environmental
food and beverages
分析物
说明
Fluorene
Di-n-butyl phthalate, analytical standard, ampule of 5000 mg
Acenaphthene-d10
Acenaphthylene
Benz[a]anthracene
Benzo[a]pyrene
Benzo[k]fluoranthene
Benzyl butyl phthalate
Bis(2-ethylhexyl) phthalate
Dibenz[a,h]anthracene
Diethyl phthalate
Indeno[1,2,3-cd]pyrene
Perylene-d12
Phenanthrene
Pyrene
Anthracene
Chrysene
Naphthalene
Acenaphthene
Benzo[b]fluoranthene
Dimethyl phthalate