推荐产品
等级
certified reference material
制造商/商品名称
Jaytee Biosciences Ltd (CRM-2001)
技术
UV/Vis spectroscopy: suitable
一般描述
CRM for UV/Visible Spectrophotometers.
These UV Stray Light Standards are designed to qualify the Stray Light of UV Spectrophotometers to the limits defined in the EP.
Developed to be used in any laboratory regardless of the regulator and are as relevant to a pharmaceutical company as to a contract lab working to ISO 17025.
All products come supplied in flame-sealed ampoules to increase the shelf life and prevent contamination, with full certification.
Traceable instrument qualification adds confidence to all data generated. In addition, your audit trail is backed up by the independent verification by UKAS to ISO 17034.
These UV Stray Light Standards are designed to qualify the Stray Light of UV Spectrophotometers to the limits defined in the EP.
Developed to be used in any laboratory regardless of the regulator and are as relevant to a pharmaceutical company as to a contract lab working to ISO 17025.
All products come supplied in flame-sealed ampoules to increase the shelf life and prevent contamination, with full certification.
Traceable instrument qualification adds confidence to all data generated. In addition, your audit trail is backed up by the independent verification by UKAS to ISO 17034.
应用
UV Stray Light Standards for use in the qualification of UV Spectrophotometers.
特点和优势
- Product is of certified reference material grade
- An ampoule of standard and an ampoule of blank are included
- CoA has the instruction on how to qualify the instrument
储存分类代码
12 - Non Combustible Liquids
WGK
WGK 2
闪点(°F)
Not applicable
闪点(°C)
Not applicable