product name
Purospher® STAR Si (5 µm), L × I.D. 250 mm x 4 mm LiChroCART® HPLC cartridge, manuCART™ holder required for use
物料
stainless steel column
质量水平
Agency
suitable for USP L3
产品线
Purospher®
特点
endcapped: no
包装
1 ea of
参数
250 bar pressure
65 °C max. temp.
技术
HPLC: suitable
LC/MS: suitable
mass spectrometry (MS): suitable (High)
色谱柱长度 × 内径
25 cm × 4 mm
表面积
330 m2/g
基质
fully porous particle
基质活性基团
Silica bonding phase
粒径
5 μm
孔径
120 Å pore size
operating pH range
2-7.5
分离技术
hydrophilic interaction (HILIC)
normal phase
储存温度
no temp limit
相关类别
一般描述
Purospher® STAR Si endcapped, 5 µm HPLC columns offers highest separation efficiency for normal-phase chromatography of low molecular weight compounds soluble in organic solvents.Purospher® STAR Si, 5 µm LiChroCART® 250-4 is a cartridge column which requires a manuCART™ cartridge column holder part number 1.51486.0001. The manuCART™ can be used to hold one cartridge column with or without direct integration of a 4-4 mm guard column. The recommended guard column is Purospher® STAR RP-8 endcapped (5 µm) LiChroCART® 4-4 part number 1.50249.0001. LiChroCART® cartridge columns mounted with a manuCART™ cartridge column holder is hand-tight only and have therefore a maximum pressure stability of 250 bar. The column dimension of 250 mm length x 4 mm I.D. is the optimal choice for the HPLC separation of complex samples allowing significant solvent savings and improved sensitivity.Learn more about Purospher® STAR HPLC and UHPLC ColumnsLearn more about the Supelco HPLC Column portfolio and Method development:Brochure: HPLC and UHPLC Column Selection GuideBrochure: A Practical Guide to HPLC - Method Development Guide
应用
Intended to be used for the separation of aflatoxins, carotenoids, oils, fat-soluble vitamins, lipids, and fatty acids.
特点和优势
- Ultra-pure silica (99.999%) for excellent peak symmetry
- High separation efficiency
- Reproducible results from run-to-run and batch-to-batch
- Use with Normal-phase or HILIC conditions (needs to be flushed to HILIC mobile phase)
- Good suitability for LC-MS applications
分析说明
Pressure: ≤ 65 bar
Theoretical Plates (N/m) (2-Nitrotoluene): ≥ 65000
Symmetry (2-Nitrotoluene): 0.8 - 1.4
Capacity factor (2-Nitrotoluene): 0.5 - 2.0
Theoretical Plates (N/m) (2-Nitrotoluene): ≥ 65000
Symmetry (2-Nitrotoluene): 0.8 - 1.4
Capacity factor (2-Nitrotoluene): 0.5 - 2.0
法律信息
LICHROCART is a registered trademark of Merck KGaA, Darmstadt, Germany
MANUCART is a trademark of Merck KGaA, Darmstadt, Germany
Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany
储存分类代码
10-13 - German Storage Class 10 to 13
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门