推荐产品
产品名称
Purospher®STAR RP-18 封端(5 µm), L × I.D. 125 mm x 2 mm LiChroCART™ HPLC cartridge, manuCART™ holder required for use
物料
stainless steel column
质量水平
Agency
suitable for USP L1
产品线
Purospher®
特点
endcapped
包装
1 ea of
标记范围
17% Carbon loading
参数
250 bar pressure
65 °C max. temp.
技术
HPLC: suitable
LC/MS: suitable
mass spectrometry (MS): suitable (High)
色谱柱长度 × 内径
12.5 cm × 2 mm
表面积
330 m2/g
基质
fully porous particle
基质活性基团
C18 bonding phase
粒径
5 μm
孔径
120 Å pore size
工作pH范围
1.5-10.5
分离技术
reversed phase
储存温度
no temp limit
正在寻找类似产品? 访问 产品对比指南
一般描述
Purospher® STAR RP-18 endcapped, 5 µm HPLC columns are designed for universal use and allow the separation of basic, neutral, and metal chelating compounds with simple mobile phases and excellent peak symmetry. They are based on ultra-pure fully porous silica particles (Type B). These columns offer an outstanding stability from pH 1.5 to 10.5 over a wide temperature range, and suitability for up to 100% aqueous mobile phases. Plus, they demonstrate best all-round retention characteristics, as proven by the Tanaka test. Thanks to these features, Purospher® STAR RP-18 endcapped, 5 µm HPLC columns allow maximum flexibility to use the best chromatographic conditions for any separation including reversed phase.Purospher® STAR RP-18 endcapped, 5 µm LiChroCART™ 125-2 is a cartridge column which requires a manuCART™ cartridge column holder part number 1.51486.0001. The manuCART™ can be used to hold one cartridge column with or without direct integration of a guard column. LiChroCART™ cartridge columns mounted with a manuCART™ cartridge column holder are hand-tight only and have therefore a maximum pressure stability of 250 bar. For higher pressure stability requirements Purospher® STAR RP-18 endcapped, 5 µm Hibar® RT 125-2 HPLC columns are available part number 1.50596.0001.The column dimension of 125 mm length x 2 mm I.D. is the optimal choice for HPLC separation of complex samples at lower flow rate, allowing significant solvent savings and improved sensitivityLearn more about Purospher® STAR HPLC and UHPLC ColumnsLearn more about the Supelco HPLC Column portfolio and Method development:Brochure: HPLC and UHPLC Column Selection GuideBrochure: A Practical Guide to HPLC - Method Development Guide
应用
Intended to be used in the separation of aflatoxins, aldehydes, alkaloids, antibiotics, catecholamines, explosives, oligonucleotides, esters, fat soluble vitamins, lipids, fatty acids, flavonoids, ketones, nitrosamines, nucleosides, PAHs, PCBs, peptides, pesticides, phenols, phthalates, preservatives, steroids, metabolized steroids, sulfonamides, sweeteners, and water soluble vitamins.
特点和优势
- Ultra-pure silica (99.999%) for excellent peak symmetry
- High separation efficiency
- Reproducible results from run-to-run and batch-to-batch
- Best all-round performance according to Tanaka
- Outstanding pH stability from pH 1.5 to 10.5
- No phase collapse when using highly aqueous mobile phases
- Good suitability for LC-MS applications
分析说明
Pressure: ≤ 65 bar
Theoretical Plates (N/m) (hexylbenzene): ≥ 70000
Symmetry (hexylbenzene): 0.8 - 1.4
Capacity factor (hexylbenzene): ≥ 8.0
Theoretical Plates (N/m) (hexylbenzene): ≥ 70000
Symmetry (hexylbenzene): 0.8 - 1.4
Capacity factor (hexylbenzene): ≥ 8.0
法律信息
Hibar is a registered trademark of Merck KGaA, Darmstadt, Germany
LICHROCART is a trademark of Merck KGaA, Darmstadt, Germany
MANUCART is a trademark of Merck KGaA, Darmstadt, Germany
Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门