G1007494
UHPLC Analysis of Nucleosides on Purospher® STAR RP-18 endcapped
application for HPLC
Recommended Products
technique(s)
HPLC: suitable
test parameters
column: Hibar Purospher® STAR RP-18 endcapped, 150 x 4.6 mm, 3 μm (1.50470.0001)
mobile phase: [A] 30 mM ammonium phosphate, pH 5.3 with ammonium hydroxide in water; [B] Methanol; 97.5 [A] / 2.5 [B]
flow rate: 1.0 mL/min
pressure: 3060 psi 211 Bar
column temp.: 25 °C
detector: UV 254
injection: 5 μL
sample: 5 μg/mL of 12 nucleosides in 98:2 30mM ammonium phosphate pH 5.3 with ammonium hydroxide in water: methanol
suitability
application for HPLC
application(s)
life science and biopharma
Analysis Note