Skip to Content
Merck
CN
All Photos(1)

Documents

G005615

HPLC Analysis of Vitamin D2 and D3 on Ascentis® Express C18

application for HPLC

Sign Into View Organizational & Contract Pricing

technique(s)

HPLC: suitable

test parameters

column: Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)
mobile phase: acetonitrile
flow rate: 0.15 mL/min
column temp.: 35 °C
detector: UV, 290 nm
injection: 10 μL
sample: 10 mg/L each (vitamins D2 and D3) in methanol

suitability

application for HPLC

application(s)

clinical
vitamins, nutraceuticals, and natural products

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    Ergocalciferol (D2), analytical standard
    Cholecalciferol (D3), analytical standard

standard

Product No.
Description
Pricing

Certificates of Analysis (COA)

Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service