Skip to Content
Merck
CN

HPLC Analysis of Ritalinic Acid and Ritalin (Methylphenidate) Enantiomers on Astec® CHIROBIOTIC® V2 with T2 Guard Column

HPLC Analysis of Ritalinic Acid and Ritalin (Methylphenidate) Enantiomers on Astec® CHIROBIOTIC® V2 with T2 Guard Column application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIROBIOTIC® V2 Chiral HPLC Column

5 μm particle size, L × I.D. 15 cm × 4.6 mm

standard

Product No.
Description
Pricing

Methylphenidate hydrochloride

Ritalinic acid

99% (CP)

CONDITIONS

column

CHIROBIOTIC T2 guard column, 2 cm x 4 mm I.D., 5 μm particles (16100AST (requires holder 21150AST))

column

CHIROBIOTIC V2, 15 cm x 4.6 mm I.D., 5 μm particles (15023AST)

mobile phase

[A] 20 mM ammonium acetate, pH 4.1; [B] methanol; (7:93, A:B)

flow rate

1 mL/min

column temp.

25 °C

detector

UV, 230 nm

injection

2 μL

sample

1 mg/mL in methanol

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC