- An erythrosin B-based "turn on" fluorescent sensor for detecting perfluorooctane sulfonate and perfluorooctanoic acid in environmental water samples.
An erythrosin B-based "turn on" fluorescent sensor for detecting perfluorooctane sulfonate and perfluorooctanoic acid in environmental water samples.
Because of the serious harm to animals and the environment associated with perfluorooctane sulfonate (PFOS) and perfluorooctanoic acid (PFOA), a rapid, sensitive and low-cost method for detecting PFOS and PFOA is of great importance. In this paper, a novel sensing method has been proposed for the highly sensitive detection of PFOS and PFOA in environmental water samples based on the "turn-on" switch of erythrosine B (EB)-hexadecyltrimethylammonium bromide (CTAB) system. In pH 8.55 Britton-Robinson (BR) buffer, EB can react with CTAB by electrostatic attraction, resulting in a strong fluorescence quenching of EB. With a subsequent addition of the CTAB, a red-shift occurred (11 nm), followed by a significant increase in fluorescence at high surfactant concentrations. It was found that PFOS and PFOA can obviously enhance fluorescence intensity of EB-CTAB system. The enhanced fluorescence intensity is proportional to the concentration of PFOS and PFOA in the range of 0.05-10 μM with detection limit of 12.8 nM and 11.8 nM (3σ), respectively. The presented assay has been successfully applied to sensing PFOS and PFOA in real water samples with RSD ≤ 4.3% and 2.9%, respectively.