Z802654
Hellma® liquid calibration standards set
UV60, UV600, UV14, for testing the photometric accuracy
Sign Into View Organizational & Contract Pricing
All Photos(1)
About This Item
Recommended Products
grade
certified reference material
manufacturer/tradename
Hellma 667-UV305
application(s)
food and beverages
petroleum
pharmaceutical
plastics
format
neat
Looking for similar products? Visit Product Comparison Guide
General description
Consisting of :
60 mg Potassium dichromate in perchloric acid (667-UV60, wavelength 235; 257; 313; 350 nm)
600 mg Potassium dichromate in perchloric acid (667-UV600, wavelength 430 nm)
Perchloric acid reference filter (667-UV14, wavelength 235; 257; 313; 350 nm)
Potassium dichromate dissolved in perchloric acid is a highly suitable method to check for photometric accuracy (absorbance) in the UV region. Potassium dichromate gives a spectral scan containing characteristic peak maxima at 257 nm and 350 nm and minima at 235 nm and 313 nm. The potassium dichromate solution is sealed in an airtight sealed quartz precision cell with 10 mm light path. The absorbance values of the filter are determined at the wavelengths 235 nm, 257 nm, 313 nm, and 350 nm and are specified in the certificate. It should be noted that the measurements are carried out with the filter 667-UV14 (0.01 N perchloric acid) in the reference beam path. Measured values of the reference filter are separately displayed in the certificate.
60 mg Potassium dichromate in perchloric acid (667-UV60, wavelength 235; 257; 313; 350 nm)
600 mg Potassium dichromate in perchloric acid (667-UV600, wavelength 430 nm)
Perchloric acid reference filter (667-UV14, wavelength 235; 257; 313; 350 nm)
Potassium dichromate dissolved in perchloric acid is a highly suitable method to check for photometric accuracy (absorbance) in the UV region. Potassium dichromate gives a spectral scan containing characteristic peak maxima at 257 nm and 350 nm and minima at 235 nm and 313 nm. The potassium dichromate solution is sealed in an airtight sealed quartz precision cell with 10 mm light path. The absorbance values of the filter are determined at the wavelengths 235 nm, 257 nm, 313 nm, and 350 nm and are specified in the certificate. It should be noted that the measurements are carried out with the filter 667-UV14 (0.01 N perchloric acid) in the reference beam path. Measured values of the reference filter are separately displayed in the certificate.
Application
Hellma® liquid calibration standards set for checking the photometric accuracy of spectrophotometers in accordance with European Pharmacopoeia (Ph. Eur.) and U.S. Pharmacopeia (USP) 857.
Features and Benefits
- Accredited to DAkkS Deutsche Akkreditierungsstelle (National accreditation body for the Federal Republic of Germany)
- Traceable to NIST (National Institute of Standards and Technology)
- Assured Quality of the measurement results
- Proven accuracy and reliability
Legal Information
Hellma is a registered trademark of Hellma GmbH
Storage Class Code
10 - Combustible liquids
Flash Point(F)
Not applicable
Flash Point(C)
Not applicable
Regulatory Information
危险化学品
Choose from one of the most recent versions:
Certificates of Analysis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service