Z802530
Hellma® liquid calibration standards set
UV6, UV9, for testing the spectral resolution acc. to Ph. Eur.
About This Item
Recommended Products
manufacturer/tradename
Hellma 667.200-UV
Looking for similar products? Visit Product Comparison Guide
General description
Consisting of :
Toluene in n-Hexane (667-UV6, scan 265 - 270 nm)
n-Hexane reference filter (667-UV9, scan 265 - 270 nm)
To check the resolution power of a spectrophotometer, the ratio of the peak maximum at λmax (269 nm) to the absorbance of the minimum at λmin (266 nm) is calculated. This gives a measure of the spectral bandwidth of the instrument.
Determination of the resolution is carried out against the 667-UV9 hexane blank reference filter, which is supplied with every filter set. Measurement is the same for all bandwidths. The wavelength range between 265 nm and 270 nm is scanned and the peak maxima and minima are determined. Then the peak/trough ratio is calculated and compared to the reference values.
Legal Information
Signal Word
Danger
Hazard Statements
Precautionary Statements
Hazard Classifications
Aquatic Chronic 2 - Asp. Tox. 1 - Flam. Liq. 2 - Repr. 2 - Skin Irrit. 2 - STOT RE 2 Inhalation - STOT SE 3
Target Organs
Central nervous system, Nervous system
Storage Class Code
3 - Flammable liquids
Flash Point(F)
-14.8 °F - closed cup
Flash Point(C)
-26.0 °C - closed cup
Regulatory Information
Choose from one of the most recent versions:
Certificates of Analysis (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documents section.
If you need assistance, please contact Customer Support.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service