Skip to Content
Merck
CN
All Photos(1)

Documents

MPGL10ZW1

Milli-Q

Millipak® Filter

Removes bacteria from pure water before entering the analyzer., For use with AFS® systems

Synonym(s):

Millipak® 100 Bacteria Filter

Sign Into View Organizational & Contract Pricing


About This Item

UNSPSC Code:
41104212
eCl@ss:
33050190
NACRES:
JA.14

material

PVDF membrane

Quality Level

sterility

non-sterile

feature

hydrophilic

packaging

pkg of 1 unit

manufacturer/tradename

Millipak®

L

13 cm (5.12 in.)

inlet connection diam.

1/4 in.

outlet connection diam.

1/4 in.

impurities

<0.5  EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

gravimetric extractables

≤2.5 mg/device

matrix

Durapore®

pore size

0.22 μm pore size

fitting

NPT inlet connection
NPT outlet connection

compatibility

for use with AFS®

shipped in

ambient

Looking for similar products? Visit Product Comparison Guide

General description

Millipak® membrane products are designed for routine filtration applications in the laboratory. Millipak® filters are best-in-class proven products, utilized in lab water purification systems that provide particulate-free and bacteria-free purified water to benefit the stringent requirements for pure water.
Millipak® 100 Bacteria Filter removes bacteria from pure water before entering the analyzer.

Application

Provides bacteria-free and particulate-free water suitable for laboratory use, including clinical analyzers.
For use with AFS® water purification systems.

Features and Benefits

  • Hydrophilic Durapore® PVDF membranes for high flow rates and throughput.
  • Low protein binding membrane yields high sample recovery with minimal loss of valuable product ensuring high-value filtration.
  • Reduces the risk of microbial contamination.
  • Ideally suited for high value-added applications.
  • Device Configuration: Capsule

Components

0.22 μm Durapore® filter membrane for bacteria removal

Other Notes

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Directions for Use

  • Organism Retention: Microorganisms
  • Mode of Action: Filtering
  • Application: General laboratory analysis
  • Intended Use: Water purification
  • Instructions for Use: This item provides water filtered through 0.22 μm sterilizing-grade membranes. Refer to system equipment user guide section "Using the System" (available on USB key in the system box).
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Preparation Note

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Analysis Note

Bacterial Retention
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
Gravimetric Extractables: after 24 hours in water at controlled room temperature

Legal Information

AFS is a registered trademark of Merck KGaA, Darmstadt, Germany
ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIPAK is a registered trademark of Merck KGaA, Darmstadt, Germany

Certificates of Analysis (COA)

Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

  1. How frequently do the various consumables need to be changed?

    The frequency that consumables must be changed depends on the system's usage (treated water volume) and on the consumables themselves (time of usage). In order to ensure the quality of water and your results, the consumables should be replaced when the system alerts you.

  2. What is the consumables delivery time?

    Refer to the consumable's Product page for a first estimation of its expected delivery time and contact customer service for more information.

  3. How are consumables changed?

    To help you to change your system's consumables, several video tutorials are available on our YouTube channel: Milli-Q®System Care Tutorials. For high-flow systems, please keep in mind that all consumable changes must be approved by a 3-digit code that only the person in charge of the system knows.

  4. What is the shelf life for Filter?

    The shelf life is 2 years for all consumables kept in the recommended storage conditions (temperatures from 5 to 30°C; humidity between 10 and 80%). Contact your local representative for more information.

  5. Can I arrange to have my consumables automatically ordered and delivered once a year?

    It's possible to place only one consumable order per year with a specific contract called "Consumable Supply Agreement (CSA)" For further information, contact your local representative or sales contact.

  6. Where can I find data sheets, Certificates of Origin, Certificates of Analysis, and Certificates of Quality?

    Documents and certificates are available in the "Documentation" section on this page. Do not hesitate to contact us if you require further information.

  7. What steps are you taking towards improved sustainability?

    Sustainability is at the heart of our innovations. From reducing water and energy consumption, to minimizing plastic and chemical waste, our dedicated water system engineers are constantly working on new ways to reduce the ecological footprint of our products to better support the sustainability goals we’re all striving towards. For more information, we invite you to review the following brochure and infographic which highlight our efforts and achievements: link. Do not hesitate to contact your local representative for further information.

  8. How can I know which consumables can be used with my system?

    To know which consumables can be used on your system, you can see in the user manual of you system or see it on the following webpage of consumables. All the consumables for each system are indicated.

  9. If I have a final filter that I use only occasionally (e.g., Biopak®) and have another filter (e.g. Millipak®) that I most of the time, can I keep the 'occasional-use' filter for use again at a later time?

    It is not possible to re-use a filter that has been used and then removed because air intake during replacement creates an airlock and blocks water passing through the filter. If you regularly need two different filters, we recommend using two POD dispensers.

  10. Can consumables be opened and refilled?

    Unfortunately, this is not possible because of safety issues and the need to guarantee the quality of consumables and their interior purification components. Our consumables are manufactured in an ISO 9001-certified facility France in agreement with high quality standards.

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.