推荐产品
无菌性
sterile-filtered
质量水平
表单
liquid
保质期
Expiry date on the label.
杂质
endotoxin, tested
密度
1.083
应用
hematology
histology
储存温度
2-8°C
一般描述
应用
特点和优势
- 可使活细胞极佳恢复
- 允许选择分离不同种类的血细胞
- 防止细胞变形
- 保持细胞活力
- 最小化外来细胞干扰
- 各批次性能可重复
- 经过过滤杀菌和内毒素检测
- 在 2-8°C 避光保护条件下可稳定保存至少2年
原理
法律信息
警示用语:
Warning
危险声明
危险分类
Eye Irrit. 2 - Met. Corr. 1 - Skin Irrit. 2
储存分类代码
8A - Combustible corrosive hazardous materials
WGK
WGK 3
闪点(°F)
Not applicable
闪点(°C)
Not applicable
个人防护装备
Eyeshields, Faceshields, Gloves, type ABEK (EN14387) respirator filter
Which document(s) contains shelf-life or expiration date information for a given product?
If available for a given product, the recommended re-test date or the expiration date can be found on the Certificate of Analysis.
How do I get lot-specific information or a Certificate of Analysis?
The lot specific COA document can be found by entering the lot number above under the "Documents" section.
How do I find price and availability?
There are several ways to find pricing and availability for our products. Once you log onto our website, you will find the price and availability displayed on the product detail page. You can contact any of our Customer Sales and Service offices to receive a quote. USA customers: 1-800-325-3010 or view local office numbers.
What is the Department of Transportation shipping information for this product?
Transportation information can be found in Section 14 of the product's (M)SDS.To access the shipping information for this material, use the link on the product detail page for the product.
Are there limitations when attempting to use Product 10831, Histopaque®-1083 with the various Accuspin products?
Yes. There are some limitations. Small amounts of blood cannot be processed in the Accuspin tubes. At least 3 mls of whole peripheral blood is expected. Without sufficient red cell numbers, the mononuclear band fails to adequately clear the barrier or frit found within the Accuspin tubes.
I find mononuclear cells separate well from human blood over Product 10831, Histopaque®-1083, but the cell recovery from rats and mice is lower. Why?
The most common cause for separation problems with Histopaque® 1083 is the blood source. Blood from rats and mice is harder to obtain than blood from a human. Most users experiencing problems are collecting blood with a syringe. It is relatively uncommon for users to call in about poor separation when using vacuum tubes. Humanely sacrifice the animal if possible and immediately collect the blood from the aorta or vena cava. Collecting blood from a severed head is never acceptable. Blood obtained from bleeding the tail is also generally unacceptable. Tissue fluids and other extraneous debris will often result in an activation of the coagulation cascase. Blood drawn into a vacuum tube is the best source of blood. See the product insert for Histopaque® 1083 for the amount of anticoagulant to add to blood collected with a syringe.
Can Product 10831, Histopaque®-1083, be used with human blood?
Cells can be recovered from human blood using Histopaque® 1083. However Sigma sells Histopaque® 1077 specifically for use with human blood. Histopaque® 1077 is intended for In Vitro Diagnostic Use. Histopaque® 1083 is intended for research use only with animal blood.
Is there any data for expected percent recovery with rat or mouse blood when using Product 10831, Histopaque®-1083?
No. Histopaque® 1083 is a research use product and no studies have been performed to determine percent recovery.
My question is not addressed here, how can I contact Technical Service for assistance?
Ask a Scientist here.
商品
Centrifugation enables the separation of particles by sedimentation. Learn how to separate particles using a centrifuge and how to use Stokes' law to calculate the velocity of sedimentation.
我们提供完整的密度梯度介质产品线,用于分离或提取白细胞、病毒、DNA、RNA、细胞器以及许多其他应用。我们的产品线包括用于在研究实验室中分离白细胞的Histopaque®碘化梯度介质,用于哺乳动物细胞分离的多糖介质,用于哺乳动物细胞和细胞器分离的胶体二氧化硅介质,以及用于DNA、病毒和蛋白质分离的无机盐介质。
实验方案
This troubleshooting guide addresses the most common sources of error observed when using Histopaque but is not meant to be a comprehensive list.
本疑难排查指南针对使用 Histopaque 时观察到的最常见错误来源提出解决方案,但这并不代表本指南是全面无遗的。
Ficoll 400® is a non-ionic polymer with high hydrophilicity and water solubility, suitable for various biological applications.
Ficoll 400®是由蔗糖和环氧氯丙烷共聚而成的一种高度分支化的聚合物。Ficoll 400®是完全非离子性的。由于其含有丰富的羟基,Ficoll 400®具有很强的亲水性和水溶性。
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系技术服务部门