Photometric Determination of Total Phosphorus in Plant Materials by Phosphorus Molybdenum Blue Method Subsequent to an Acid Extraction

Abstract

The protocol details the photometric determination of total phosphorus in plant materials using the phosphorus molybdenum blue method. Following digestion transforming all phosphorus to phosphate, a phosphomolybdate complex is formed. Spectroquant® test kits and spectrophotometers are employed to deliver precise results, aiding in the analysis of plant nutrient levels and the effectiveness of soil, manure, or fertilizers used in cultivation.

Section Overview:

three models of Spectroquant® spectrophotometers: Prove 100 plus, Prove 300 plus, and Prove 600 plus. Each device has a sleek, modern design with a rectangular body. The largest model in the foreground, Prove 600 plus, is green with a black top and features a touchscreen interface displaying measurement data, as well as a slot for sample insertion. The top right corner of the device has a yellow "M" logo and "600+" printed in bold yellow on the front corner. Two smaller models, Prove 100 plus and Prove 300 plus, are placed in the background. Prove 100 plus is light green with a black top, while Prove 300 plus is yellow with a black top. Both also feature touchscreens and the "M" logo.


Spectroquant® spectrophotometers - Prove 100 plus, Prove 300 plus, & Prove 600 plus

Introduction

Phosphorus, along with nitrogen and potassium is an important macronutrient for the growth and development of plants. Specifically, phosphorus plays an important role in the development of the root system, and in the development of the nutrient uptake system by the plant. Plants obtain phosphorus from soil mainly as phosphates. It has been observed that irrespective of plant species, the general phosphorus content should be 0.05-0.5% of a plant’s dry weight1. Phosphorus deficiency can lead to underdevelopment and hampered growth of a plant. Even when a plant grows in soil containing a sufficient concentration of phosphorus, it can be phosphorus-deficient when that soil has phosphorus in the form of inorganic phosphates, which have low solubility and high sorption capacity. It is therefore important to analyze total phosphorus in plant materials, to study the quality of the soil, manure, or fertilizer used for cultivation.  

This application note details a relatively inexpensive estimation of phosphorus by the molybdate method in plant materials. The described determination of phosphorus (total) is based on the formation of the phosphomolybdate complex with digestion. 

Reagents, Instruments and Materials

Phosphate Test Kits/Reagents

For the measurement, one of the following Spectroquant® test kits is necessary:

  • Spectroquant® Phosphate Cell Test (1.14543)
  • Spectroquant® Phosphate Cell Test (1.14729)
  • Spectroquant® Phosphate Test (1.14848)

Instrument(s) & Devices

For the measurement, one of the following Spectroquant® photometers is necessary:

  • Spectroquant® VIS Spectrophotometer Prove 100 Plus (1.73026)
  • Spectroquant® UV/VIS Spectrophotometer Prove 300 Plus (1.73027)
  • Spectroquant® UV/VIS Spectrophotometer Prove 600 Plus (1.73028)
  • Spectroquant® Colorimeter Move 100 (1.73632)

Also, legacy systems

  • Spectroquant® Spectrophotometer Prove 100/300/600
  • Spectroquant® Photometer NOVA 30/60/60A

Software for Data transfer

  • Optional Spectroquant® Prove Connect to LIMS software package (Y.11086) to transfer your data into an existing LIMS system.

Instrument Accessories

  • Rectangular cells 10 mm (1.14946) and/or
  • Rectangular cells 20 mm (1.14947) and/or
  • Rectangular cells 50 mm (1.14944)

Other Reagents and Accessories

  • Hydrochloric acid 37 % for analysis (1.00317)
  • Nitric acid 65 % for analysis (1.00456)
  • Water for analysis (1.16754)
  • Empty cells 16 mm with screw caps (1.14724)

Experimental Procedure

Sample Preparation

  • In a well-covered glass beaker dissolve about 100 mg of the sample in 3 mL hydrochloric acid (37%) and 3 mL nitric acid (65%) in the steam bath.
  • When completely dissolved, evaporate the acids, to a well-dried, acid-free, crystalline solid residue.
  • After taking up a few mL of water for analysis, transfer the solution to a 1000 mL volumetric flask, fill it to the mark with water for analysis, and mix well. 

Analysis

Determine with one of the above-mentioned test kits following the instructions under “Procedure” (no additional digestion needed).

Calculation

Phosphorus (total) content in % = (analysis value in mg/L P x 5) / (weight of sample in mg)

See more applications for photometry at Protocols and Application Notes

Related Products
Product NumberProduct NameProduct DescriptionPricing
19349TraceCERT® Multiresidue Pesticide Standard Mix #1in acetonitrile, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
29985TraceCERT® Multiresidue Pesticide Standard Mix #2in acetonitrile, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
30176TraceCERT® Multiresidue Pesticide Standard Mix #3in acetonitrile, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
36082TraceCERT® Multiresidue Pesticide Standard Mix #4in acetonitrile, TraceCERT®, (Manufactured by: Sigma-Aldrich Production GmbH, Switzerland)
30639TraceCERT® Multiresidue Pesticide Standard Mix #5in acetonitrile, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
36471TraceCERT® Multiresidue Pesticide Standard Mix #6in acetonitrile, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
PHR3706除虫菊提取物certified reference material, pharmaceutical secondary standard
43065Spinetoram solution100 μg/mL in acetonitrile, certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland
52604-UDiscovery ® DSC-18 SPE 小柱bed wt. 500 mg, volume 6 mL, pk of 30
565301-UAscentis® RP-Amide HPLC 色谱柱3 μm particle size, L × I.D. 10 cm × 2.1 mm
53500-UAscentis® Express保护柱套for use with Ascentis Express Guard Columns, pk of 1
53514-UAscentis® Express RP-Amide,2.7 μm保护柱芯2.7 μm particle size, L × I.D. 5 mm × 2.1 mm, pkg of 3 ea
T2318Greiner 离心管centrifuge tube, 50 mL, 30 x 115 mm, conical (V) bottom, w/ graduations, I.D. field
T1818Greiner 离心管centrifuge tube, 15 mL, 17x120 mm, conical (V) bottom, w/ graduation, I.D. field, w/ rack
Z742705BenchMixer XLQ QuEChERs Shaker/VortexerAC/DC input 115 V AC, US 2-pin plug
Z627577BRAND® seripettor®瓶口分液器volume range 2.5 - 25 mL , Adjustable in 0.5 mL increments
29653-UCertified Vial Kit, Low Adsorption (LA), 2 mL, pk of 100volume 2 mL, amber glass vial (with marking spot), natural PTFE/silicone septa, thread for 9 mm
803410Low Dead Volume Precolumn Filter with 0.5 μm Stainless Steel Fritpkg of 1 ea
803411Replacement 0.5 μm stainless steel frits for 2 mm I.D. TSKgel® columnspkg of 10 ea
BR705872BRAND® Transferpette® S 移液器tip volume (2-20 μL), single-channel, adjustable volume, mechanical, DE-M

References

1.
Wieczorek D, Żyszka-Haberecht B, Kafka A, Lipok J. Determination of phosphorus compounds in plant tissues: from colourimetry to advanced instrumental analytical chemistry. Plant Methods. 18(1): https://doi.org/10.1186/s13007-022-00854-6
Sign In To Continue

To continue reading please sign in or create an account.

Don't Have An Account?