Skip to Content
Merck
CN

HPLC Analysis of mAb Charge Variants on TSKgel® CM-STAT

HPLC Analysis of mAb Charge Variants on TSKgel® CM-STAT application for HPLC

Materials

analytical column

Product No.
Description
Pricing

TSKgel® CM-STAT HPLC Column

phase carboxymethyl, L × I.D. 10 cm × 4.6 mm, 7 μm particle size

CONDITIONS

column

TSKgel CM-STAT, 10 cm x 4.6 mm I.D., 7 μm particles (821966)

mobile phase

[A] 20 mM MES buffer, pH 6.0; [B] 0.5 M NaCl in buffer A, pH 6.0

gradient

0% B (0 min), 30% B (15 min), 100% B (15 min),

0% B (17 min), 10% B (17 min), 10% B (21 min)

flow rate

1 mL/min

column temp.

ambient

detector

UV, 280 nm

injection

20 μL

sample

monoclonal antibodies (mAb A through E)

Description

Analysis Note

A TSKgel CM-STAT weak cation exchange (WCX) column was applied to separate charge variants of several monoclonal antibodies. The typical analysis time on conventional 25 cm long WCX columns of about eighty minutes could be significantly reduced when separation was performed on a 10 cm TSKgel CM-STAT column, filled with 7 μm particles. The analysis profiles for five antibodies show that high resolution analysis can be obtained in about 20 minutes analysis time.

Citation

From "Ion Exchange Chromatography for the Characterization of Biotherapeutics" in The Supelco Reporter, Vol. 29.3, page 20.

Legal Information

TSKgel is a registered trademark of Tosoh Corporation