If you decide to place an order during this period, we reserve the right to adjust the price based on the evolving situation. We understand that market changes may cause inconvenience. We will negotiate with you if there’s a significant price fluctuation due to tariff policy changes before the order’s actual delivery, and in such cases we may adjust or cancel the order as necessary.

For important updates on recent policy changes, please click here for more details.

Merck
CN

HPLC Analysis of Aflatoxins on Ascentis® Express C18 with Fluorescence Detection

HPLC Analysis of Aflatoxins on Ascentis® Express C18 with Fluorescence Detection application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Ascentis® Express C18, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

standard

Product No.
Description
Pricing

Aflatoxin Mix

certified reference material, in methanol (varied conc.), ampule of 1 × 5 mL

CONDITIONS

column

Ascentis Express C18, 15 cm x 4.6 mm I.D., 2.7 μm particles (53829-U)

mobile phase

[A] water; [B] acetonitrile; (75:25, A:B)

flow rate

2.0 mL/min

pressure

6381 psi (440 bar)

column temp.

35 °C

detector

fluorescence, ex 360 nm, em 440 nm

injection

50 μL

sample

200 μL aflatoxin standard in acetonitrile; 0.5 μg/mL B2 and G2; 2 μg/mL B1 and G1; 800 μL delivatization solution: 70% water, 20% TFA and 10% acetic acid
heat for 9 min at 65 °C

Description

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany