Skip to Content
Merck
CN

GC Analysis of (R)-(+)-1-Phenylethylamine (N-Trifluoroacetyl Derivative) on Astec® CHIRALDEX G-PN

GC Analysis of (R)-(+)-1-Phenylethylamine (N-Trifluoroacetyl Derivative) on Astec® CHIRALDEX™ G-PN suitable for GC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIRALDEX G-PN Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.12 μm

CONDITIONS

column

Astec CHIRALDEX G-PN, 30 m x 0.25 mm I.D., 0.12 μm (74033AST)

oven

130 °C

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

helium, 30 psi

injection

1 μL, 80:1 split

sample

3 mg/mL in methanol
peak 1: (R)-(+)-1-phenylethylamine (N-TFA derivative) (ee% 99.1)
peak 2: (S)-(-)-1-phenylethylamine (N-TFA derivative)

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIRALDEX is a trademark of Sigma-Aldrich Co. LLC