Skip to Content
Merck
CN

GC Analysis of α-Hexachlorocyclohexane (α-HCH) Enantiomers on Astec® CHIRALDEX G-TA

GC Analysis of α-Hexachlorocyclohexane (α-HCH) Enantiomers on Astec® CHIRALDEX™ G-TA suitable for GC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIRALDEX G-TA Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.12 μm

standard

Product No.
Description
Pricing

HCH

PESTANAL®, analytical standard, mixture of isomers (α:β:γ:δ=1:1:1:1)

α-HCH

PESTANAL®, analytical standard

CONDITIONS

column

Astec CHIRALDEX G-TA, 30 m x 0.25 mm I.D., 0.12 μm (73033AST)

oven

170 °C

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

helium, 30 psi

injection

0.2 μL, 80:1 split

sample preparation

2 mg/mL in ethanol
peaks 1 & 2: α-HCH (α-hexachlorocyclohexane) enantiomers