  • Determination of matairesinol in flax seed by HPLC with coulometric electrode array detection.

Determination of matairesinol in flax seed by HPLC with coulometric electrode array detection.

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences (2002-09-25)
Tanja Kraushofer, Gerhard Sontag
ABSTRACT

A HPLC method coupled with coulometric electrode array detection for the determination of matairesinol in flax seed is described. The defatted sample was spiked with bisphenol A (internal standard), refluxed for 75 min in a mixture of ethanol-bidistilled water-12 M hydrochloric acid (2:2:1, v/v/v) to extract matairesinol conjugates and to hydrolyze them simultaneously. The extract was diluted with mobile phase [250 ml acetonitrile-750 ml buffer (730 ml bidistilled water, 20 ml glacial acetic acid adjusted to pH 3 with 5 M sodium hydroxide)] and injected into the HPLC system. Matairesinol was separated from other compounds on a reversed-phase column (Lichrospher 60 RP-Select B, 250 x 4 mm, 5 micro m) and detected in a coulometric electrode array detector using a flow-rate of 0.8 ml/min. The potentials of the eight electrodes were set on +150, +200, +250, +300, +350, +400, +550 and +600 mV against modified palladium electrodes. The content of matairesinol determined in seven samples varies between 7 and 28.5 micro g/g. The limit of quantitation is 5 micro g/g.