  • [Method study on dehydroaetic acid in food by high performance liquid chromatography].

[Method study on dehydroaetic acid in food by high performance liquid chromatography].

Wei sheng yan jiu = Journal of hygiene research (2008-02-29)
Huizhen Zhang, Aiguo Ma, Hui Liang, Yongye Sun
ABSTRACT

To estabish a high performance liquid chromatographic (HPLC) method for the determination of dehydroaetic acid in food determinated. Separation was carried out on a column of C18 (51 microm, 250 x 4.60mm) at the temperature of 25 degrees C with methanol/0.02mol/L ammonium acetate (5:95, V/V) as mobile phase, and elution flow rate was 1.0ml/min. The detection wavelength was 306nm and the injection volume was 10 microl, quantitation was used by external standard method. The results showed a good linear relationship between the peak area and the concentration (r = 0.9997), the recoveries for Dehydroaetic acid products were 96.5% - 99.5%, and the relative standard deviation was 1.09% - 5.85% for different kinds of food. The results showed that the method may be simple, rapid, accurate and applicable for the determination of dehydroaetic acid in food.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
Dehydroacetic acid, ≥98.0% (T)