Skip to Content

很抱歉,目前系统维护中。在此期间,网站下单功能无法使用。给您带来不便,深表歉意。如需其他帮助,请联系客户联络中心:400 620 3333。

Merck
CN
HomeSolid Phase Microextraction (SPME)Standard Methods 6040D: GC Analysis of Geosmin and 2-MIB on SLB®-5ms after SPME using 50/30 µm DVB/Carboxen/PDMS Fiber

Standard Methods 6040D: GC Analysis of Geosmin and 2-MIB on SLB®-5ms after SPME using 50/30 µm DVB/Carboxen/PDMS Fiber

Standard Methods 6040D: GC Analysis of Geosmin and 2-MIB on SLB®-5ms after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber

Conditions

desorption process260 °C for 3 min.
extractionheadspace, 65 °C (30 min)
SPME fiber2 cm Metal coated with 50/30 µm DVB/Carboxen/PDMS (57914-U)
columnSLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U)
oven60 °C (2 min), 8 °C/min to 200 °C
carrier gashelium, 1 mL/min constant
sample20/10 ppt odor compounds in 25 mL water + 25% NaCl
liner0.75 mm I.D., SPME
MSD interface300 °C
scan rangeSIM, m/z = 137, 124, 95, 112

Description

CategoriesAnalytical Chromatography, SPME Applications, GC Applications, Volatiles
Featured IndustryEnvironmental
Food and Beverages
Legal InformationSLB is a registered trademark of Sigma-Aldrich Co. LLC
Other NotesComplementary products:
Syringes
Vials
Gas Purification and Gas Management
Inlet Liners, Glass Wool, and Specialized Hand Tools
Septa and Specialized Hand Tools
Column Ferrules, Nuts, and Specialized Hand Tools
suitabilityapplication for SPME, application for GC

Analyte

  • (±)-Geosmin
  • 2-Isobutyl-3-methoxypyrazine
  • 2-Isopropyl-3-methoxypyrazine
  • 2-Methylisoborneol