LC/MS/MS Analysis of Aflatoxins in Cannabis on Ascentis® Express 2.7µm Phenyl-Hexyl after Cleanup Using Supel Tox AflaZea SPE

LC/MS/MS Analysis of Aflatoxins in Cannabis on Ascentis®

Figure 1.

Conditions

Sample PreparationSolid Phase Extraction
Sample/MatrixCannabis extract spiked with aflatoxins at corresponding concentrations of 6 ppb each (for Aflatoxin B2 and G2) and 24 ppb each (for Afltoxin B1 and G1) in cannabis
SPE Tube/CartridgeSupel Tox AflaZea SPE, 6 mL (55314-U)
Sample Addition2 mL of sample extract
ElutionApply strong vacuum and collect the sample into plastic tube
Eluate Post-treatmentDilute 0.2 mL of the extract with 0.8 mL distilled water and mix well
ColumnAscentis Express Phenyl-Hexyl, 5 cm x 2.1 mm I.D., 2.7 μm particles (53334-U)
Column Temp.40 °C
Mobile Phase[A] 5 mM ammonium formate with 1% formic acid in water; [B] 5 mM ammonium formate with 1% formic acid in methanol
Gradient30% B to 60% B in 3 min; to 100% B in 2 in; held at 100% B for 2 min; to 30% B in 0.1 min; held at 30% B for 1.4 min
Flow Rate0.4 mL/min
Pressure5511 psi (380 bar)
Injection10 μL
DetectorMS, ESI(+), MRM 331.3/189.0, 329.1/243.0, 315.9/259.0, 313.1/241.0

Description

Analysis NoteThe complex composition of cannabis makes it difficult to apply sample cleanup methods commonly used for other commodities. In this application, interference removal Supel Tox AflaZea SPE was used for sample cleanup providing quick and effective sample preparation prior to mass spectrometric analysis. Sample comprised 0.5 g of finely ground cannabis mixed with 10 mL of 84:16 acetonitrile:water for 30 minutes and centrifuged. It is recommended to use silanized glass vials for this analysis.
CategoriesAnalytical Chromatography, Drugs of Abuse, Mycotoxins
Featured IndustryFood and Beverages
Vitamins, Nutraceuticals, and Natural Products
Legal InformationAscentis is a registered trademark of Sigma-Aldrich Co. LLC, Supel is a trademark of Sigma-Aldrich Co. LLC
SuitabilityApplication for LC-MS, application for SPE